Vulnerability: Hard-coded accounts in Hospira Infusion Systems

Vulnerability: Hard-coded accounts in Hospira Infusion Systems

CVE-2015-3953 · HIGH Severity

AV:N/AC:L/AU:N/C:C/I:C/A:C

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.