Unsecured Bluetooth LE Implementation in Dr Trust ECG Pen 2.00.08 Devices Allows Data Sniffing and Man-in-the-Middle Attacks

Unsecured Bluetooth LE Implementation in Dr Trust ECG Pen 2.00.08 Devices Allows Data Sniffing and Man-in-the-Middle Attacks

CVE-2020-15486 · MEDIUM Severity

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.