Vulnerability: Unauthorized Access and Impersonation in Fresenius Kabi Vigilant MasterMed v2.0.1.3

Vulnerability: Unauthorized Access and Impersonation in Fresenius Kabi Vigilant MasterMed v2.0.1.3

CVE-2021-23207 · MEDIUM Severity

CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:N

An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users.

Learn more about our Physical Security Assessment.